Please visit our website for more information www.alpine-components.co.
Page Table of contents 1. Application 23 2. Safety instructions 23 3. Errors and load restrictions 23 4. Safety regulations 23 4.1 Special conditions for use in Zone 0 areas 24 5. Safety instructions 24 6. Ex data 24 7. Technical data 24 8. Functional description / operating information 26 8.1 Operation 26 8.2 Overview 26 8.3 Display screen 27 8.4 Settings 27 8.4.
1. Application The Ex-MX2 is a contactless temperature measuring device designed for industrial use in potentially explosive areas of Zones 2 and 1 in accordance with Directive 1999/92/ EC (ATEX 137). 2. Safety instructions These operating instructions contain information and safety regulations that must be followed to ensure safe and reliable operation of the unit under the described conditions.
During use, always make sure that the leather pouch is completely closed and undamaged. - Only accessories approved by ecom instruments GmbH may be used. - Avoid using the unit in aggressive acidic or alkaline solutions. - Ensure that the unit is not taken into Zone 0 areas. 4.1 Special conditions for use in Zone 0 areas Temperature measurement with the Ex-MX2 is only allowed within Zone 0 only if the following conditions are met: 1.
60:1 at the focus point (1.15 m) and Measuring spot size (distance / Ø): 35:1 in the far field (> 10 m) Accuracy (25 °C): ± 0.75% of the measured value or ± 0.75K, whichever is the larger value (± 2K for measurement objects below -5°C) Ambient derating (at 25 °C ± 25K): <0.05K/K or <0.05%/K, whichever is the larger value Reproducibility: ± 0.5% of the measured value or ± 0.5K, whichever is the larger value Spectral sensitivity: 8 ...
8. Functional description / operating information Please take the time to carefully read through these operating instructions to ensure that you are familiar with and can use all the functions on your Ex-MX2. For your own safety and information please read through the information on the following pages. 8.1 Operation All objects with a temperature above absolute zero radiate infrared energy which spreads in all directions at the speed of light.
8.3 Display screen Continuous measurement Laser Maximum setting reached Measured temperature Bar display of the last 10 measured values Battery condition Emissivity MIN/MAX display of the last measurement 8.4 Settings System settings can be fixed and activated with the switch panel.
.4.1 Emissivity Most organic materials and lacquered and oxidised surfaces have an emissivity of 0.95. The emissivity is preset to 0.95 in the Ex-MX2 for this reason, but can be changed with the rocker switch. The determination of the temperatures of shiny or highly polished metal surfaces produces inaccurate measured values. To compensate for this, the surface to be measured can be covered with adhesive tape or painted matt black.
8.5.2 Optical diagram Ø mm measuring spot Distance m 25 0 19 1.15 29 1.50 58 2.50 72 3.00 8.5.3 Field of view Ensure that the measuring object is larger than the size of the measuring spot of the device. The smaller the measuring object the closer you must approach the object. If the measurement accuracy is crucially important, the measuring object should be at least twice as large as the size of the measuring spot. 9.
. Troubleshooting Code -O-UE-EPROM-Err CalAreaErr ProbCalErr Blank display Battery symbol is displayed Laser does not function Fault Target temperature is above the measuring range Target temperature is below the measuring range E-EPROM fault Calibration error Calibration error Weak or empty batteries Batteries are almost empty Weak or discharged batteries or ambient temperature >+45 °C Procedure Select target within the range Select target within the range Contact the manufacturer Contact the manufactu
In accordance with “Directive 2006/66/EC on batteries and accumulators and waste batteries” we are obligated to provide you with the following information: Your device contains an alkaline battery. Depleted batteries should never be disposed of along with normal or household waste. Old batteries can contain harmful substances that are hazardous to health and damaging to the environment. Please return the batteries. Return is free of charge and required by law.
If in certain countries the restriction of statutory warranty and the exclusion or restriction of incidental or consequential damages is unlawful, then it may be possible that the above restrictions and exclusions do not apply for all purchasers. If any clause in these warranty terms should be found to be invalid or unenforceable by a competent court, then such a judgement shall not affect the validity or enforceability of any other clause contained in these warranty terms.
15.
Communication Portable Lighting Mobile Computing Measurement & Calibration mobile experts Angewandte harmonisierte Standards Applied harmonized standards Standards harmonisés appliqués EN 50014:1997 +A1+A2 EN 50020:2002 DIN EN 61326:1997 EN 61010-01:1993 + A2 EN 60825-1:1994 Angewandte europäische Richtlinien Applied European directives Directives générales européennes appliquées 94/9/EG / 94/9/EC 89/336/EWG / 89/336/ECC 73/23/EWG / 73/23/ECC ecom instruments GmbH Assamstadt 08. 01.
16.
36
37
38
39
40
5. Amendment to certificate number: Device: Manufacturer: Address: TÜV 00 ATEX 1597 X Explosion-proof temperature-measuring device Type Ex-MX2 or Ex-MX4 BT ecom instruments GmbH Industriestr. 2 D-97959 Assamstadt Commission no.: 8000553725 Modifications: The explosion-proof temperature-measuring device type Ex-MX2 or Ex-MX4 BT may be produced and operated in the future in accordance with the test documents listed below.
(18) Fundamental safety and health requirements No additional requirements TÜV NORD CERT GmbH, Langemarckstrasse 20, D-45141 Essen, accredited by the Zentralstelle der Länder für Sicherheits Technik (Central Office of the Federal States for Safety Technology), ID no. 0044, legal successor of TÜV NORD CERT GmbH & Co. KG, ID no.
Kommunikation Mobile Leuchten Mobile Computing Mess- & Kalibriertechnik Communication Portable Lighting Mobile Computing Measuring & Calibration Copyright© 2010 ecom instruments GmbH. Der Inhalt dieses Dokuments darf ohne vorherige schriftliche Genehmigung durch die ecom instruments GmbH in keiner Form, weder ganz noch teilweise vervielfältigt, weitergegeben, verbreitet oder gespeichert werden. Der Inhalt dieses Dokuments wird so präsentiert, wie er aktuell vorliegt.