beeps will be heard and two (2) green LED flashes will be visible. Each push of the “-” button will produce an audible
beep and the red LED will flash once indicating one (1) increment down. When the lowest setting is reached, two (2)
audible beeps will be heard and two (2) red LED flashes will be visible. To reset a Bed Beacon’s range back to the factory
default setting, push and hold both of the “+” and “–“ buttons at the same time. Four (4) audible beeps will be heard
and the LEDs will flash in an alternating pattern between green and red. The factory default range setting for the Bed
Beacon is position 9. Replace the battery cover when the adjustment is complete.
Figure 9. Bed Beacon Range Adjustments
5.2 Bed Beacon to Test Badge Range Adjustment
2. Set and adjust the bed and mattress height to the average height for beds in the facility.
3. While holding the Test Badge 50 inches from the floor and 24 inches from the bed’s handrail center location
(side of bed) (see Figures 10 & 11), the Test Badge’s red and yellow LEDs should flash once a second, indicating
that the proper activation range and communication have been successfully established. If range is not 24”
adjust the Bed Beacon’s range buttons until the Badge activation range is 24 inches (see Figure 9).
Figure 10. Test Badge Height Is 50 Inches From the floor
Decrement “-“
Increment “+”