Patient Safeguard System® Patient Bed Installation
UltraClenz, LLC
Page 10
2. Place cable tie holders and route the ALP Sensor cable along the inside rail and tie down any excess cable (Service
Loop) as shown (see Figure 7).
Figure 7. ALP Sensor Cable Routing and Cable Tie Holder Placement
1. Place a cable tie holder and route the ALP Sensor cable as shown (see Figure 8).
Figure 8. ALP Sensor Cable Routing to Bed Beacon and Cable Tie Holder Placement
5. Adjusting the Communication Range
5.1 Range Adjustment on Bed Beacon
The Bed Beacon communication range is adjusted during installation to maximize communication with the healthcare
worker Badge. To adjust the communication range, the range buttons are located on the back of the Bed Beacon. The
Bed Beacon communication range is adjusted by repeatedly pushing either the “+” increment or the –“decrement
range buttons (see Figure 9) to achieve the desired range. The total adjustable range of the Bed Beacon is 1 to 15
increments between the lowest and the highest setting. Each push of the “+” button will produce an audible beep and
the green LED will flash once indicating one (1) increment up. When the highest setting is reached, two (2) audible
Head of the bed
Foot of the bed
Head of the bed
Foot of the bed
Tie Downs