Patient Safeguard System® Patient Bed Installation
UltraClenz, LLC
Page 10
Figure 6. AC Line Power Sensor Attached to Power Cord
Cable Tie mounting
Mount the AC Line Power Sensor on patient bed’s AC power cord using 2 cable ties (see Figure 6). Use cable tie
holders for routing the wire to the Dock, and make sure the mounting surface is clean and free of dirt and oil. A soft
cloth dampened with isopropyl alcohol works well to clean both surfaces.
5. Adjusting the Communication Range
5.1 Range Adjustment on Bed Beacon
The Bed Beacon communication range is adjusted during installation to maximize communication with the healthcare
worker Badge. To adjust the communication range, remove the battery cover located on the front of the Bed Beacon
using a #2 Phillips screwdriver. Note: the battery cover has a retaining ring to prevent the screw from being detached.
After the cover has been removed, the Bed Beacon communication range is adjusted by repeatedly pushing either the
“+” increment or the “–“decrement range buttons (see Figure 8) to achieve the desired range. The total adjustable
range of the Bed Beacon is 1 to 15 increments between the lowest and the highest setting. Each push of the “+” button
will produce an audible beep and the green LED will flash once indicating one increment up. When the highest setting is
reached, two (2) audible beeps will be heard and two (2) green LED flashes will be visible. Each push of the “-” button
will produce an audible beep and the red LED will flash once indicating one increment down. When the lowest setting is
reached, two (2) audible beeps will be heard and two (2) red LED flashes will be visible. To reset a Bed Beacon’s range
back to the factory default setting, push and hold both the “+” and the “–“ buttons down at the same time. Four (4)
audible beeps will be heard and the LEDs will flash in an alternating pattern between green and red. The factory default
range setting for the Bed Beacon is position 9. Replace the battery cover when the adjustment is complete.
Figure 7. Bed Beacon Range Adjustments
Decrement “-“
Increment “+”
AC Line Power Sensor
attached to power cord