User Manual
Table Of Contents
FAS1533 User's Guide UltraClenz, LLC
Page 5
4.2 Procedure
1. Apply power to the dispenser beacon
2. Push and release the + range switch four times
3. The blue led will flash once with each activation
4. Push and release the + range switch once (fifth position)
5. The led will flash twice indicating the range is set to maximum. If this does not
occur depress both the + and - simultaneously to set to factory default . The blue
led will flash 4 times to indicate factory default range. Test again.
6. Push and release the - range switch four times
7. The blue led will flash once with each activation
8. Push and release the - range switch once (fifth position)
9. The led will flash twice indicating the range is set to maximum. If this does not
occur depress both the + and - simultaneously to set to factory default . The blue
led will flash 4 times to indicate factory default range. Test again.
5.0 Event Activation
5.1 Procedure with a Badge
1. Apply power
2. Place a test badge 24" or 61 cm or less near the beacon
3. Trigger a dispense event on the phono jack
4. The blue led's will alternate twice indicating a badge was detected
5. The event will be transmitted with the badge address
5.2 Procedure without a Badge
1. Apply power
2. Trigger a dispense event on the phono jack
3. The blue led's not light
4. The event will be transmitted without a badge address
Fig.3 - Range Switches