User Manual

DFU Document Title Here
UltraClenz, LLC
Page 5
Green & Red LEDs
The green and red LEDs provide visual feedback to the healthcare worker (See Figure.1 item 4).
GREEN: The single green led flash indicates a door entry or exit with compliance.
RED: Indicates a wash compliance warning.
Audible Alert
The audible alert makes a beeping sound that is used to give feedback to the healthcare worker (See Figure.1 item 5).
Bypass Button
The bypass mode must be enabled in the PSS Door Monitor configuration software, see section 3.0 PSS Door Monitor
Configuration (Figure 4) for this function to work. The override button is used to stop the PSS Door Monitor from
sensing for a programmed time period (See Figure 1 item 6). The override mode can be used by housekeeping staff to
prevent multiple non-compliance events while performing routine room maintenance.
The Override Time can be set as follows:
Momentarily depress the override button once for a 30-second default countdown. The override button grace period
can be changed from the PSS Door Monitor configuration software (See Figure 4).
Adjustable Sensor
The Sensor can be adjusted to select the correct PIR angle. The sensor adjustment is done at the time of installation to
optimize door entry and exit sensing (See Figure 1 item 7).
Front Cover
Removable front cover and screw assembly (See Figure. 1 item 8).
1.2 Component Descriptions - Back View (See Figure. 2)
Dispenser Connector
The dispenser plugs into this connector of the PSS Door Monitor (See Figure. 2 item 9). The PSS Door Monitor is
attached to a dispenser via an interface connector and counts all patient room entries and exits with dispense
(compliant event), without dispense (non-compliant event) and total room entries and exits.
2. PSS Door Monitor Led Functionality
2.1 Hand Hygiene Compliance Warning (Red LED)
The hand hygiene compliance LED warning will flash and the audible alarm will beep to alert the HCW entering or
exiting the patient room. The alert will continue until the HCW has sanitized or the preset pass-through timer has timed
out (see PSS Door Monitor Configuration Software). If the pass-through timer reaches the set terminal count, the event
will be recorded and the red indicator LED will stop flashing.
2.2 Hand Hygiene Compliant - Enter or Exit Area (Green LED)
When a HCW activates the dispenser, the hand hygiene compliant LED will flash and the audible alarm will beep once.
The event will be counted only if the HCW enters or exits the patient room. The PSS Door Monitor will not record an
event if dispense activation occurs without a door entry, however the beeps will still occur.
2.3 Low Battery Indicator (Red LED)
The PSS Door Monitor will produce a single red LED flash every 3 seconds indicating a low battery condition.
2.4 Dead battery Indicator (Red LED)
The PSS Door Monitor will produce a double red LED flash every 3 seconds indicating a dead battery condition.