Installation Instructions
7
To verify which mode the Bed Beacon is in, press and hold both range buttons down. Immediately, the Beacon
will flash three (3) times the green LED. With each flash, there will be an audible alert (beep). If the range buttons
are released during or immediately after the three green flashes, the bed beacon is set to the default range (10 =
maximum). One (1) second later, the Bed Beacon will flash red and beep once. The single red LED flash and
audible alert corresponds to the Bed Beacon being set for mode number 1: bed mode. If both range buttons
continue to be held down, the bed beacon will continue to cycle through the operation modes, at a rate of 1 every
500 ms. The number of red LED flashes and audible beeps corresponds to the operation mode (see list below.)
In order to select the mode, release the range buttons after the desired mode is reached.
An alternative method of setting the bed mode is with the Ecolab Beacon Installation Tool (9250-3074). For
more information about how the Beacon Installation Tool can change the Beacon mode, please read the
document entitled, “Ecolab™ Hand Hygiene Program Compliance Monitoring System Beacon Installation Tool
User’s Guide”.
Operation Modes:
1 = Bed Mode using Internal Antenna Only
2 = Bed Mode using External Antenna Only
3 = Bed Mode using Internal and External Antenna (toggle between internal and external antenna)
4 = Stretcher Mode using Internal Antenna Only
5 = Stretcher Mode using External Antenna Only
6 = Stretcher Mode using Internal and External Antenna (toggle between internal and external antenna)
Note:
Bed Modes (1-3) - AC Power Cord Sensor must detect AC power to broadcast the 125kHz signal.
Stretcher Modes (4-6) – the 125kHz signal is always broadcast. No AC Power Cord Sensor required.
The default operation mode is 1. To select a different mode, press and hold both range buttons simultaneously. The LEDs
will flash and the piezo will beep to indicate the mode.
***
3 green flashes and beeps
–
default NFC range, which is 10 (maximum)
*
1 red flash and beep
–
operation mode 1
**
2 red flashes and beeps
–
operation mode 2
***
3 red flashes and beeps
–
operation mode 3
****
4
red flashes and beeps
–
operation mode 4
*****
5 red flashes and beeps
–
operation mode 5
******
6 red flashes and beeps
–
operation mode 6
.