Installation Instructions

35
Figure 1. Bed Beacon Installation Location on the Bed
1.3 Step 3: Installing the AC Line Power Sensor on the Bed’s AC Power Cord
The AC Line Power Sensor (see Figure 2) is mounted onto the bed’s black AC power cord located
underneath the bed. Use the supplied cable ties to secure the AC Line Power Sensor to the bed’s power
cord as shown. The Bed Beacon field is only enabled when the Sensor detects AC power on the bed’s line
cord after it is plugged into an AC outlet. The Bed Beacon field will be disabled while the bed is unplugged.
The Bed Beacon will send heartbeats in both the enabled and disabled modes.
1.4 Step 4: Adjusting the Bed Beacon to Proper Field Range
1. Connect the AC Line Power Sensor to the Bed Beacon via any RJ11 jacks as shown in Figure 3.
AC Line Power
Sensor attached to
the bed’s power cord
Figure 2.
The
AC Line Power Sensor
Installation Location on the Bed
AC Line Power
Sensor Cable RJ11
Connectors
Bed Beacon