8. Troubleshooting
8.1 No Bed Beacon Heartbeats on the Dashboard
If the Bed Beacon has stopped transmitting heartbeats to the Dashboard, there are three possible causes:
1. The batteries have died. Place a new Duracell Coppertop alkaline D set of batteries in the Bed Beacon
and test the Bed Beacon’s other function, communications with a Badge as described in section 6.3. If
the unit tests well, then give the Bed Beacon a couple of hours to transmit a few heartbeats to the
Dashboard.
2. If new batteries do not fix the problem, there may be a problem with the Bed Beacon hardware. Remove
and replace the Bed Beacon. Refer to section 7.2 – Replacing an Existing Bed Beacon – Swapping
Metadata -- for instructions on how to replace/update device information on the Dashboard.
3. If replacing the Bed Beacon does not remedy the problem, there may be a problem with the Network
Contact your Ecolab representative for further assistance.
8.2 The Bed Installation Does Not Communicate with a Badge
Verify that the bed’s AC power cord is plugged into an active power source and that the AC Line Power
Sensor is connected to the bed’s power cord.
Note: If any of the components of the Bed Installation are replaced, the bed’s range will have to be re-
checked and readjusted as needed.