Installation Instructions
15
Figure 12. Test Badge Height Is 50 Inches From the floor
Figure 13. Test Badge Activation Distance from Bed’s Handrail Center is 18” (at edge for stretcher installations)
6.3 Ecolab Dashboard Test with the Network Badge and Bed Beacon
If needed, please refer to “Ecolab™ Hand Hygiene Program Compliance Monitoring System Dashboard
Directions for Use” for further information on how to use the Dashboard.
1. Using a HCW Badge, verify that the Bed Beacon and Badge activate at 18 x 50 inches. Also verify that
the Bed Beacon’s device address appears on the Dashboard.
2. Using a web browser, type the facility’s Dashboard URL and you should see login screen shown in
Figure 14.
3. Type in the correct email address and password to login to the Dashboard.
Figure 14. Dashboard Entry Screen
4. When access is granted you should see the screen shown in Figure 15.