Temperature & Pressure Data Logger EBI 12
for routine control and process validation in steam sterilizers, washer disinfectors
RED LINE
The EBI 12 data logger is distinguished by
its fast response time and simple connectivity.
Hose connection and M10x1 internal thread
External temperature sensor Ø1.95 mm, rigid, axial
EBI 12-TP237
Type Description Part-No.
EBI 12-TP237 L = 20 mm, Hose connection 1340-6658
Technical Data:
0 °C … +150 °C
0 °C … +150 °C
± 0.1 °C (0 °C … +120 °C)
± 0.05 °C (+120 °C … +140 °C)
± 0.1 °C (+140 °C … +150 °C)
± 20 mbar
0.01 °C
1 mbar
2x 33,440 measurements
PT1000, class A
Piezoresistive pressure sensor
250 ms … 24 hours
Yes
-20 °C … +70 °C
endless measurement
Measure upon starttime
Start immediately until end of
memory
Start- / Stop measurement
Lithium, 3.6 V,
User-replaceable
(Ø x H) 48 mm x 24 mm
app. 90 g
316 L stainless steel / PEEK
IP68
0 °C, +60 °C and +134 °C
100 mbar and 3,100 mbar (at +25 °C)
3,100 mbar (at +134 °C)
Standards and Directives:
EN 285
Sterilization, steam sterilizers,
medical equipment
EN 13060
small steam sterilizers
ISO 17665
Sterilization of health care products
Moist heat
ISO 15883
Washer Disinfectors
2014/53/EU
Radio Equipment Directive - RED

Summary of content (1 pages)