Datasheet
Xylem Analytics Germany Sales GmbH & Co.KG, ebro -Peringerstr. 10 – D-85055 Ingolstadt, Germany
Tel +49 841 954780 – Fax +49 841 95478-80
Website: www.ebro.com – E-Mail: ebro@xylem.com
ebro® is a registered trademark of Xylem Inc. or one of its subsidiaries. Technical changes reserved.
© 2014 Xylem Analytics Germany Sales GmbH & Co. KG. March 2022
Temperature & Pressure Data Logger EBI 12
for routine control and process validation in steam sterilizers, washer disinfectors
RED LINE
The EBI 12 data logger is distinguished by
its fast response time and simple connectivity.
• Hose connection and M10x1 internal thread
• External temperature sensor Ø1.95 mm, rigid, axial
EBI 12-TP237
Type Description Part-No.
EBI 12-TP237 L = 20 mm, Hose connection 1340-6658
Technical Data:
Operating temperature
0 °C … +150 °C
Measurement range
0 °C … +150 °C
Accuracy: temperature
± 0.1 °C (0 °C … +120 °C)
± 0.05 °C (+120 °C … +140 °C)
± 0.1 °C (+140 °C … +150 °C)
Accuracy: pressure
± 20 mbar
Resolution: temperature
0.01 °C
Resolution: pressure
1 mbar
Data memory
2x 33,440 measurements
Sensor: temperature
PT1000, class A
Sensor: pressure
Piezoresistive pressure sensor
Measurement interval
250 ms … 24 hours
Real-time data
transmission
Yes
Storage temperature
-20 °C … +70 °C
Measurement modes
• endless measurement
• Measure upon starttime
• Start immediately until end of
memory
• Start- / Stop measurement
Battery
Lithium, 3.6 V,
User-replaceable
Dimensions
(Ø x H) 48 mm x 24 mm
Weight
app. 90 g
Housing material
316 L stainless steel / PEEK
Protection class
IP68
Factory Calibration
Certificate
0 °C, +60 °C and +134 °C
100 mbar and 3,100 mbar (at +25 °C)
3,100 mbar (at +134 °C)
Standards and Directives:
• EN 285
Sterilization, steam sterilizers,
medical equipment
• EN 13060
small steam sterilizers
• ISO 17665
Sterilization of health care products
— Moist heat
• ISO 15883
Washer Disinfectors
• 2014/53/EU
Radio Equipment Directive - RED