Do not block the inlet or outlet areas of the enclosure and keep at
least one foot of clearance around all sides of the bedside controller
to prevent overheating.
Do not place Ebb bedside controller on the floor where it can be
bumped or where someone is likely to trip over the power cord.
Not for use in an oxygen-rich or flammable anesthetic environment.
To avoid electrical shock, always unplug the power cord from the
wall before cleaning the bedside controller. Do not drop or place the
bedside controller in any fluids or allow fluids to enter vent openings.
If you notice any of the following, disconnect power from the bedside
controller and contact Ebb Customer Care.
– Changes in the performance of Ebb
– Unusual or harsh sounds
– If the bedside controller has been dropped or mishandle
– If the enclosure of the bedside controller is broken.
There are no user-serviceable components inside the device. Do not
attempt to open the bedside controller.
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches) to any part of Ebb device,
including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
WARNING Warnings alert the user about situations that could result in injury or death if not avoided.
!
WARNING Warnings alert the user about situations that could result in injury or death if not avoided.
!
• Read the entire manual before use.
Some individuals may experience skin redness on the forehead
aer wearing the forehead pad. If this persists more than 12 hours,
contact Ebb Customer Care.
The headband must be worn in the proper orientation to prevent
entanglement and strangulation.
The headband must be routed above the pillow to prevent
entanglement and strangulation. Improper routing could cause a
kink and result in loss of therapy.
Patients with cold sensitivity reactions such as Raynaud’s Disease
should consult their doctor before using cooling therapies in order
to avoid cold sensitivity reactions that may include cold sensations,
color changes in the skin in response to cold, a numb prickly feeling
or stinging pain upon warming.
If you have a skin condition on your forehead that may make
wearing the headband uncomfortable, please check with your
doctor before use.
Do not use Ebb if the package or any part is damaged. Contact Ebb
Customer Care.
Route the electrical cord to prevent a tripping hazard or being
entangled with furniture.
Isolation from Supply Mains, Electrical Hazards: Do not position Ebb
so it is difficult to unplug the power cord from the electrical outlet.
Do not use extension cords.
Use only a power supply provided by Ebb Therapeutics. Use of any
other power supply may result in increased emissions or decreased
protection of Ebb which could impact the performance of the unit.
Keep the power supply away from hot surfaces.
Ebb is isolated from the mains by means of the supplied medical
grade power supply. Do not use a power supply other than the one
supplied with the system.
Ebb should not be used close to or stacked with other electrical
equipment.
!