- Change the cantilever if no improvement can be seen after these proce-
dures.
Set point drift
When part of the scan line in the Line graph is drawn red, the tip has moved
to its maximum or minimum Z-position. This should also be visible in the
range display on the right hand side of the graph. The tip will move to its
maximum Z-position when it has lost contact with the sample:
- In the Static Force based modes, increase the force Set point, in the
Dynamic Force based modes, reduce the amplitude Set point.
If this does not help:
- Click in the Approach panel.
No connection to microscope
This error message appears when the easyScan 2 software is waiting for an
answer from the scan electronics. Most likely, the easyScan 2 Controller is
not connected to the mains power, or not turned on. In this case the status
lights on the top of the electronics are not on.
Check the connection and the power switch.
USB Port error