Operating instructions

PROBLEMS AND SOLUTIONS
66
Problems and Solutions
The problems described here can occur during normal operation of the
microscope. If the suggested course of action does not solve the problem,
or the problem is not described here, refer to the section Nanosurf support
(p.68).
Probe Status light blinks red
This light will blink red when an insufficient amount of light reaches the
cantilever deflection detector. This can be either be due to the DropStop
still being closed or to a misaligned cantilever chip.
- Check that the cantilever is lying perfectly in the alignment chip.
- Remove the cantilever using the cantilever insertion tool (figure Contents
of the AFM Tool set (p.14), 5). Blow away dust from the alignment chip
and the cantilever’s back side using dry, oil free air. Remount the canti-
lever.
For further information, open the Scan Head Diagnosis dialog via the
menu ‘Options/Check Scan Head...’.
Image quality suddenly deteriorates
Tip picks up material
When a scan line suddenly starts reproducing badly, the tip may have
picked up some particles:
- Continue measuring for a while (1-2 images), the tip may eventually lose
these particles again.
- Retract the cantilever by clicking then perform a new approach.
- Try to induce changes in the tip’s end: While measuring increase the
force ‘SetPoint’, or decrease the amplitude ‘SetPoint’ in the Feedback
panel. Restore the set point to its old value when the contrast improves,
or if nothing happens after scanning several lines.