Operating instructions

IMPROVING MEASUREMENT QUALITY
52
Ideally, the XY-plane of the scanner has already been correctly aligned with
the sample plane using the three Levelling screws on the Scan head. How-
ever, this alignment can not be performed once the automatic approach has
been done, because this will damage the tip. During measurement, the
measurement plane should therefore be adjusted electronically using the
parameters ‘X-Slope’ and ‘Y-Slope’.
To find the correct values, use the following procedure:
- The slope correction is only available with the Standard level and the
Advanced level user interface mode. Open the User interface dialog via
the menu ‘Options/Config user interface, and select one of these modes.
- Measure the slope in the Line graph using the angle tool (see Software
Reference). Use a single click to measure the angle relative to the X*-Axis.
- Alter the value of ‘X-Slope’ in the Imaging options of the Imaging panel
until the x-axis of the scan line lies parallel to the X-axis of the sample.
Displays after adjusting the slopes