Operating instructions

FINISHING
47
For an in depth introduction to the Nanosurf Report software, refer to the
Introduction section of the Nanosurf Report online help.
Finishing
Once you are done measuring:
- Click to stop measuring.
- Open the positioning window.
- Click .
- Retract the cantilever to a safe distance from the sample by clicking
until the tip-sample distance is at least as large as shown in
figure View of the cantilever after manual approach (p.41).
Turning off the instrument
- Verify that you have saved all measurements that you would like to keep.
- Exit the easyScan 2 software after having stored all desired measurement
documents. If you exit the program whilst having some unsaved measure-
ments, you will be asked to save them now.
- Turn off the power switch.
- Store the Scan head in the Scan head case.
Important!
Always store the scan head with an (old) cantilever installed. This will
prevent dust from gathering on the alignment chip, and protect the align-
ment chip against damage.