Operating instructions

INSTALLING THE CANTILEVER
29
operating mode; the more flexible, long CONTR cantilever is generally
used for the static operating mode.
When you change to a different cantilever type:
- Select the cantilever type using the ‘Mounted cantilever’ drop-down
menu in the Operating mode panel.
Presently, only cantilevers from the companies NanoWorld and NanoSen-
sors can be used. The various usable types are listed in section Compatible
cantilevers (p.11).
Inserting the cantilever in the Scan head
Important!
Nothing should ever touch the cantilever, especially the tip end!
The Cantilever Holder Spring (see figure Cantilever deflection detection
system) is very delicate and should not be bent!
Always close the DropStop before handling the cantilever. Otherwise, the
cantilever may fall into the scan head. This may cause malfunction of the
microscope, particularly the scanner.
If a cantilever has dropped into the scan head, and the microscope is mal-
functioning, contact your local support. Never open the scan head,
because this is likely to cause damage to the scanner and the laser beam
deflection detection system.
To insert the cantilever:
- Remove the Cantilever Insertion tool from the DropStop
Cantilever deflection detection system
Sample illumination
Photodetector
Laser
Cantilever
holder spring
Cantilever
Alignment chip
Hole for cantilever
insertion tool