Operating instructions
INSTALLING THE CANTILEVER
29
operating mode; the more flexible, long CONTR cantilever is generally
used for the static operating mode.
When you change to a different cantilever type:
- Select the cantilever type using the ‘Mounted cantilever’ drop-down
menu in the Operating mode panel.
Presently, only cantilevers from the companies NanoWorld and NanoSen-
sors can be used. The various usable types are listed in section Compatible
cantilevers (p.11).
Inserting the cantilever in the Scan head
Important!
• Nothing should ever touch the cantilever, especially the tip end!
• The Cantilever Holder Spring (see figure Cantilever deflection detection
system) is very delicate and should not be bent!
• Always close the DropStop before handling the cantilever. Otherwise, the
cantilever may fall into the scan head. This may cause malfunction of the
microscope, particularly the scanner.
• If a cantilever has dropped into the scan head, and the microscope is mal-
functioning, contact your local support. Never open the scan head,
because this is likely to cause damage to the scanner and the laser beam
deflection detection system.
To insert the cantilever:
- Remove the Cantilever Insertion tool from the DropStop
Cantilever deflection detection system
Sample illumination
Photodetector
Laser
Cantilever
holder spring
Cantilever
Alignment chip
Hole for cantilever
insertion tool