Operating instructions

THE EASYSCAN 2 AFM
16
The Controller
Status lights
All status lights on top of the Controller will light up for one second when
the power is turned on.
The Probe Status light
Indicates the status of the Z-Feedback loop. It can indicate the following
statuses:
red The scanner is in its upper limit position. This occurs
when the tip-sample interaction is stronger than the
set point for some time. There is danger of damaging
the tip due to too high an interaction.
orange/yellow The scanner is in its lower limit position. This occurs
when the tip-sample interaction is weaker than the set
The easyScan 2 Controller
Video Out
connector
(optional)
Probe Status light
Scan head
cable connector
Video In
connector
(optional)
Signal Out
connector
(optional)
Signal In
connector
(optional)
Scan Head lights
Module lights
Power
switch
USB power
light
Mains power
connector
USB active
light
USB input
(from PC)
Controller
Serial number
USB outputs
(to dongle)
S/N: 23-05-001