Manual
USER GUIDE 
76 
Vario Averager Seconds:  This setting controls the averaging period of the variometer, in 10ths of a second 
units. For faster response, lower periods are recommended, but in rougher air longer periods may be required 
to avoid false alarms. 
Vario Minimum Sinkrate (Deadband): This setting controls the minimum sinkrate. Normally, this would 
be set to the standard sinkrate of your model, i.e., the rate at which it descends in smooth air with no thermals. 
When you descend at a rate greater than this sinkrate, the variometer will start to sound 
Vario Minimum Climbrate (Deadband): This setting controls the minimum climbrate. When you ascend at 
a rate greater than this climbrate, the variometer will start to sound, 
Set Vario Audible Tone Gain: This setting changes the amount the variometer pitch changes as the climbrate 
increases or decreases. If you are not hearing enough change, increase this parameter, and vice versa for too 
much frequency change. 
Total Energy Comp Percent: This adjusts the scaling of the Total Energy Compensation calculation. 100% 
means that Kinetic Energy is converted into Potential Energy directly, i.e., any change in airspeed is assumed to 
directly negate any change in altitude. Less than 100% reduces the degree which airspeed change affects Total 
Energy, and increasing it above 100% increases the amount. 
Vario Off when Motor Running: This option disables the variometer tones when the motor is running, based 
on the amperage draw of the motor. This is useful for “motor 
gliders.” The variometer is off when motor current is more 
than approximately 1.5 amps. 
7.7 Vector Calibration 
The Vector is factory calibrated, and normally no further 
calibration is necessary. However, user calibration of some 
sensors can be performed. Calibration options are located on 
the “Calibration and Sensor Setup” menu. 
7.7.1 Electrical Calibration 
The voltage and amperage (current) readings for your motor 
pack, and the voltage readings for your transmitter pack, 
camera pack and receiver pack (if used) can be calibrated. 
These calibrations are performed in the “Electrical Calibration” 
menu. For your convenience, all voltages and currents are 
displayed in real time on the electrical calibration page, so you 
can compare the Vector readings with the readings of your 
voltage and current meter as you change the settings. 
Voltage Calibration: To calibrate voltages, change the “Voltage 
Factor” for the appropriate voltage. Increasing the factor 
increases the reported voltage, and vice versa. 
Current Calibration: Calibrating the current sensor typically involves just increasing or decreasing the 
“Current Sensor Factor” as appropriate.  
One additional calibration setting for the current sensor is the “Zero Offset”.  The zero offset is chosen to match 
the lowest level of current flow that the Vector’s current sensor can detect. Typically, this is about 300 mA (0.3 
amps). 










