10.4 Pin specications
Pin Name Type* Description Level Electrical specication
1 VSEL AI Set value: voltage
0…10V or 0...5V correspond
to 0..100% of U
Nom
Accuracy < 0,2%
Impedance Ri >100K
2 CSEL AI Set value: current
0…10V or 0...5V correspond
to 0..100% of I
Nom
3 N.C. Not connected
4 DGND POT
Reference potential for
digital control signals
For +Vcc, control and status signals
5 REMOTE DI
Toggle between internal
or external control
External = LOW***, U
Low
<1V
Internal = HIGH, U
High
> 4V
Internal = open
U range = 0 …30V
I
Max
= +1mA at 5V
Sender: Open collector against DGND
6 OT DO Overtemperature error
OT/PF = HIGH, U
High
> 4V
no error = LOW***, U
Low
<1V
Quasi open collector with pull-up to Vcc **
At 5V at the output there will be max.+1mA
I
Max
= -10mA at U
CE
= 0.3V
U
Max
= 0...30V
Short-circuit-proof against DGND
7 N.C. Not connected
8 PSEL AI Set value: power
0…10V or 0...5V correspond
to 0..100% of P
Nom
Accuracy < 0.5%
Impedance Ri >100k
9 VMON AO Actual value: voltage
0…10V or 0...5V correspond
to 0..100% of U
Nom
Accuracy < 0.2% at I
Max
= +2mA
Short-circuit-proof against AGND
10 CMON AO Actual voltage: current
0…10V or 0...5V correspond
to 0..100% of I
Nom
11 AGND POT
Reference potential for
analogue signals
For -SEL, -MON signals
12 N.C. Not connected
13 REM-SB DI Output off
off = LOW***, U
Low
<1V
on = HIGH, U
High
> 4V
on = OPEN
U range = 0…30V
I
Max
= +1mA at 5V
Sender: Open collector against DGND
14 OVP DO Overvoltage error
OVP = HIGH, U
High
> 4V
no OVP = LOW***, U
Low
<1V
Quasi open collector with pull-up to Vcc **
At 5V at the output there will be max.+1mA
I
Max
= -10mA at U
ce
= 0.3V
U
Max
= 0...30V
Short-circuit-proof against DGND
15 CV DO
Indication of voltage
regulation active
CV = LOW***, U
Low
<1V
CC = HIGH, U
High
>4V
* AI = Analogue input, AO = Analogue output, DI = Digital input, DO = Digital output, POT = Potential
** Internal pull-up voltage 13...14V *** Default setting, can be modied in the setup menu
10.5 Example applications
10.5.1 General
Note: Switching the DC output on when it is initially off does not
work if remote control is not active.
Note: A digital output of a PLC or any other controlling applica-
tion may not be able to pull inputs like REMOTE or REM-SB
to LOW. Refer to the technical specications of the particular
hardware for details.
Principle view of input REM-SB and REMOTE:
+
10.5.2 Output off/on
Note: Pin REM-SB is not depending on remote control being
active and can switch off the DC output at any time resp. inhibit
to switch it on, except the device is in LOCAL mode. During
LOCAL mode the device can only be controlled manually. See
section 6.9 about LOCAL.
Switching the DC output of the device off is either done with
LOW or HIGH level on the pin 13, REM-SB (input). This depends
on the settings related to the analogue interface in the device
setup (see section 10.3).
Attention! This pin is internally pulled up to HIGH. When selec-
ting HIGH in the setup menu for this pin, the function „output
off“ would be permanent and would require to pull the pin to
LOW rst, in order to be able to manually or digitally control
the DC output.
Condition LOW can be achieved by using a low-resistive contact
like a switch, open collector transistor or relay.
Operating the device