Operating the device
Attention! Never connect any ground of the analogue
interface to minus (negative) or plus (positive) output
of the device!
This will eliminate the galvanic isolation and possibly
put dangerous potential to the analogue interface
and thus also to the control application (eg. a PLC),
especially when operating the DC output on trans-
formerless inverters.
Figure 10
10.3 Settings in the device setup
+Analogue interface +
The setup menu gives access to some settings for the built-in
analogue interface:
Analogue voltage Default: 0...10V
= { 0...5V | 0...10V } Selects the voltage range for the
analogue in- and outputs, which cor-
responds to 0...100% set/actual value
REMOTE /5 Default: LOW
REM-SB /13 Default: LOW
= { LOW | HIGH} Denes, if the digital inputs will react
their dedicated function with either LOW
or HIGH level. For the levels of LOW
and HIGH see section 10.4.
Note: both pins are internally pulled-up to HIGH. It means,
that if setting HIGH is selected and nothing is connected to
the analogue interface, pin REMOTE would instantly switch to
permanent remote control and pin REM-SB would switch the
output permanently off.
OVP /14 Default: LOW
OT /6 Default: LOW
CV /6 Default: LOW
= { LOW | HIGH} Denes, if the digital outputs will report
their dedicated status with either LOW
or HIGH level. For the levels of LOW
and HIGH see section 10.4.