Instruction Manual
PSI 8000 3U HS PV Series
Date: 10-28-2011
About the power supply
The state of the power output is displayed in the bottom right
corner of the display.
{ON,OFF} State of the power output
The presently active control mode is displayed to the right of
the related actual values. The output values are limited by the
active control mode:
- limited by the voltage set value
(= Constant Voltage)
- limited by the power set values
(= Constant Power)
- limited by the set value of current
(= Constant Current)
- limited by the set value for internal resistance
(optional at U/I/R mode), indicated next to the
actual voltage
(= Constant Resistance)
Additionally to the state of the output an alarm, a warning or a
signal can be displayed:
Alarm Example: = Overtemperature
Warnings Example: = Overvoltage
Signals Example: = Overcurrent
The location from where the unit is currently controlled is dis-
played below the output state. This location is absolute, which
means that you cannot control the unit from elsewhere without
changing the location.
local Control only possible at the unit
remote Remote control via communication interfaces
(IF-C1, IF-R1, IF-U1 etc.)
extern Remote control via built-in or optional, analogue
interface
6.4 Switching the power output on
By pressing the ON button the power supply output
is switched on, as long as it is not overridden by the input pin
„REM-SB“ (13) of the built-in analogue interface or optional ana-
logue interface card IF-A1, because the pin has higher priority.
If so and when trying to switch the output on by the button, the
display will indicate the status text „auto ON“, noticing the user
that the output will switch on as soon as the override from the
pin is removed.
The output state is indicated in the display with ON.
Note: in local state (see section 6.9), the pin REM-SB of the
analogue interface (internal or external) is inoperative.
The display shows the current state with „ON“.
The OFF button switches the power supply output
off. This state is displayed with OFF.
6.5 Adjusting set values
As long as extern or remote is not displayed, the set
values for voltage, current or power can be set manually.
The way of set value submission is selected in the device se-
tup at Accept set value. The setup is accessed with button
-> Prole -> General settings -> Control panel.
See „7.4. Conguring the control panel“ for details.
Direct setting of the set values
Using the rotary knobs directly sets the set values.
The left rotary knob adjusts the voltage. The set value of the
voltage is displayed invertedly while it is selected and adjusted.
The right rotary knob either sets the set value for the current, for
the power or internal resistance (optional, unlockable, with U/I/R
mode chosen). The selected set value is displayed invertedly.
With the SELECT keys
the set value for the power, with
the set value for the internal resistance or with
the set value for the current is selected.
The maximum adjustable power can also be limited.
Set values are submitted
Alternatively to the direct adjustment of set values you can
choose to set the set values only after submitting them with
the RETURN button. See section „7.4. Conguring the control
panel“ for details. The set values can still be changed with the
rotary knobs, but are not set to the output as long as they’re
not submitted. While the set value is unchanged, only its unit
is displayed invertedly. If the set value is changed it is also
displayed invertedly.
The SELECT keys switch from current adjustment to power
adjustment for the right rotary knob. The chosen set values are
not submitted to and set by the power supply until then.
Pressing the RETURN button submits the set values.
Pressing the ESC button discards the new set values
and the old set values are displayed again.
Note: the adjustment of the resistance set value is only
accessible after the optional „internal resistance control“
is unlocked (see section 7.9).
The resistance set value is adjustable from 0 up to
20* Unom/Inom. Means, for example, at a device with Unom
= 1000V and Inom = 30A it can be adjusted to a maximum
of 666,6Ω.