PS 2000 B Single Series
General display ranges:
Value Unit Range Description
Actual voltage V 0.3-115% U
Nom
Actual value of DC output voltage
Set value of voltage V 0-100% U
Nom
Set value for limiting the DC output voltage
Actual current A 0.3-100% I
Nom
Actual value of DC output current
Set value of current A 0-100% I
Nom
Set value for limiting the DC output current
Limit settings A, V 0-100% I
Nom
Limit the set value ranges of voltage and current
Protection settings A, V 0-110% I
Nom
/ U
Nom
OCP (overcurrent) and OVP (overvoltage)
• Status display (lower part)
This area displays various status texts and symbols:
Display Description
The HMI is locked
The HMI is unlocked
Remote The device is under remote control via USB
Alarm Alarm condition which has not been acknowledged or still exists.
1.9.4.2 Rotary knobs
As long as the device is in manual operation the two rotary knobs are used to adjust set values as well as setting up
parameters in the menu. For a detailed description of the individual functions see section «3.4. Manual operation».
1.9.4.3 Button function of the rotary knobs
The rotary knobs also have a pushbutton function which is used anywhere during value adjustment to shift the
cursor as shown:
1.9.4.4 Resolution of the displayed values
Inthedisplay,setvaluescanbeadjustedwithaxedstepwidth.Thenumberofdecimalplacesdependsonthe
device model. The values have 3 or 4 digits.
Adjustment resolution and number of digits of set values in the display:
Voltage, OVP Current, OCP
Nominal
Digits
Step width Nominal
Digits
Step width
42 V 4 0.01 V 3 A / 5 A / 6A 3 0.01 A
84 V 4 0.01 V 10 A / 20 A 4 0.01 A