PS 3000 C Series
3.4.3.1 Menu “Settings”
This is main menu for all settings related to the general operation of the device and of the interface(s).
Sub menu Description
Output Settings Allows for adjustment of set values related to the DC output, alternatively to the
handling in the main screen of the display
Protection Settings Allows for adjustment of protection thresholds (here: OVP, OCP, OPP) related to
the DC output. Also see section „3.3. Alarm conditions“
Limit Settings Allows for adjustment of adjustment limits for set values. Also see section „3.4.4.
Adjustment limits“
General Settings Settings for the operation of the device and its interface(s). Details below
Reset Device When selecting “Yes”andconrmingwith“Enter” button, it will initiate a reset of
allsettings(HMI,proleetc.)tofactorydefault,asshowninthemenustructure
diagrams on the previous pages
3.4.3.2 Menu “General Settings”
Setting P. Description
Allow remote control 1 Selection “NO” means that the device cannot be remotely controlled over either
the digital or analog interfaces. If remote control is not allowed, the status will be
shown as “Local” in the status area on the main display. Also see section 1.9.5.1.
DC output after power ON 2 Determines the condition of the DC output after power-up.
• OFF=DCoutputisalwaysoafterswitchingonthedevice.
• Restore = DC output condition will be restored to the condition prior to switch
o.
DC output after PF alarm 2 Determines how the DC output shall react after a power fail (PF) alarm has
occurred:
• OFF=DCoutputwillbeswitchedoandremainuntiluseraction
• AUTO = DC output will switch on again after the PF alarm cause is gone and
if it was switched on before the alarm occurred
DC output after remote 2 Determines the condition of the DC output after leaving remote control either
manually or by command.
• OFF=DCoutputwillbealwaysowhenswitchingfromremotetomanual
• AUTO = DC output will keep the last condition
Analog Rem-SB action 3 This parameter is only displayed if the optional Analog/USB interface is installed.
Selects the action on the DC output that is initiated when changing the level of
analog input “Rem-SB”:
• DC OFF=thepincanonlybeusedtoswitchtheDCoutputo
• DC ON/OFF=thepincanbeusedtoswitchtheDCoutputoandonagain,
ifithasbeenswitchedonbeforeatleastfromadierentcontrollocation
Analog interface range 3 This parameter is only displayed if the optional Analog/USB interface is installed.
Selects the voltage range for the analog set values, monitoring outputs and
reference voltage output.
• 0...5 V = Range is 0...100% set /actual values, reference voltage 5 V
• 0...10 V = Range is 0...100% set /actual values, reference voltage 10 V
See also section „3.5.4. Remote control via the analog interface (AI)“
Analog interface Rem-SB 4 This parameter is only displayed if the optional Analog/USB interface is installed.
Selects how the input pin “Rem-SB” of the analog interface shall be working
regarding levels and logic:
• Normal = Levels and function as described in the table in 3.5.4.4
• Inverted = Levels and function will be inverted
Also see „3.5.4.7. Application examples“