3.8.3 Manual operation
3.8.3.1 Accessing and conguring the sequence generator
Via the touchscreen the sequence generator (SEQ) can be called up, congured
and controlled. Selection and conguration are only possible when the input is
switched off.
► How to access the SEQ
1. Make sure the DC input of the load module is switched off.
2. Tap the touch area on the main screen and in the menu tap on
and then .
3. The rst screen of the SEQ will always be the sequence point selection, as
shown in the example to the right.
► How to congure a sequence point
1. If there is one or several sequence points to congure, select the rst one in
the screen “Seq. point select” by using the rotary knob and start editing it
by tapping on touch area “Edit”. The edit screen for the selected sequence
point will be loaded (see example to the right).
2. Adjust the values for U, I and P, as well as the time, with the rotary knob
as usual. Direct input is not available here. Selecting a different value is by
tapping on it. The time value denes how long the settings for U, I and P
are valid when processing the particular sequence point.
3. Once you ar nished with setting a point, you may chose to save the changes and return to the select se-
quence point screen by tapping “SAVE” or if you are going to edit subsequent points (next or previous), use
the two touch areas “Prev.seq” and “Next seq.”.
Following adjustment ranges apply for the parameters of all sequence points:
Parameter Adjustment range Description
U 0...U
Nom
Voltage set value of the sequence point over the given time
I 0...I
Nom
Current set value of the sequence point over the given time
P 0...P
Nom
Power set value of the sequence point over the given time
Time 1 ms ... 36000 s Period of the sequence point in seconds
The adjustment limits (“Limits”) do not apply here. The time of all sequence points sums up to the total run time of one
sequence cycle. With this, a minimum time of 1 ms (1 sequence) and a maximum time of 1000 h (100 sequences.
► How to congure the sequence
1. In the screen “Seq. point select” (see above) tap on touch area “NEXT”.
This will open screen “Sequence control” (see example to the right).
2. Select the block of sequence points to process for the next run by setting
the rst (Start seq.) and last (End seq.) point.
3. Also set the number of Seq. cycles (1-999 or innite).
4. Conrm your settings with touch area “NEXT” or abort with “ESC”.
With tapping on “NEXT”, the device will nally load the sequence, switch to the actual sequence generator screen
SEQ. The sequence run can now be started.
If any of the sequence points in the block has not explicitly been set while conguring the points,
it will either have default values (U = 0, I = 0, P = 320 W, t = 0.3 s) or the ones which have been
set the last time.