EL 9000 T Series
3.2.2 Current regulation / constant current / current limitation
Current regulation is also known as current limitation or constant current mode (CC) and is fundamental to the
normal operation of an electronic load. The DC input current is held at a predetermined level by varying the internal
resistanceaccordingtoOhm’slawR=U/Isuchthat,basedontheinputvoltage,aconstantcurrentows.Once
the current has reached the adjusted value, the device automatically switches to constant current mode. However,
if the power consumption reaches the adjusted power level, the device will automatically switch to power limitation
and adjust the input current according to I
MAX
= P
SET
/ U
IN
, even if the maximum current set value is higher. The
current set value, as determined by the user, is always and only an upper limit.
While the DC input is switched on and constant current mode is active, the condition “CC mode active” will be
shown on the graphics display by the abbreviation CC, as well it will be passed as a signal to the analog interface
and stored as internal status which can be read via digital interface.
3.2.3 Resistance regulation / constant resistance
Inside electronic loads, whose operating principle is based on a variable internal resistance, constant resistance
mode(CR)isalmostanaturalcharacteristic.Theloadattemptstosettheinternalresistancetotheuserdened
value by determining the input current depending on the input voltage according to Ohm’s law I
IN
= U
IN
/ R
SET
.
The internal resistance is naturally limited between almost zero and maximum (resolution of current regulation
tooinaccurate).Astheinternalresistancecannothaveavalueofzero,thelowerlimitisdenedtoanachievable
minimum. This ensures that the electronic load, at very low input voltages, can consume a high input current from
the source, up to the maximum.
While the DC input is switched on and constant resistance mode is active, the condition “CR mode active” will
be shown on the graphics display by the abbreviation CR, as well it will be stored as internal status which can be
read via digital interface.
3.2.4 Power regulation / constant power / power limitation
Power regulation, also known as power limitation or constant power (CP), keeps
theDCinputpowerofthedeviceattheadjustedvalue,sothatthecurrentowing
from the source, together with the input voltage, achieves the desired value. Power
limitation then limits the input current according to I
IN
= P
SET
/ U
IN
as long as the
power source is able to provide this power.
Power limitation operates according to the auto-range principle such that at lower
inputvoltageshighercurrentcanowandviceversa,inordertomaintainconstant
power within the range P
N
(see diagram to the right).
While the DC input is switched on and constant power operation is active, the con-
dition “CP mode active” will be shown on the graphic display by the abbreviation
CP, as well it will be stored as internal status which can be read via digital interface.
Constant power operation impacts the internal set current value. This means that the maximum set current may
notbereachableifthesetpowervalueaccordingtoI=P/Usetsalowercurrent.Theuserdenedanddisplayed
set current value is always the upper limit only.
3.2.4.1 Temperature dependent derating
This series consists of conventional electronic loads which convert the consumed electrical energy into heat and
dissipate it. In order to avoid overheating, the device will automatically reduce, i.e. derate the actual input power
when heating up. It means, at a cold start it can take the peak power (see technical specs) for a certain time before
it starts reducing.
This derating is depending on the ambient temperature. It means, that at 10°C the load can take the peak power
for a much longer time than at 20°C ambient temperature or higher. Disregarding the ambient temperature, the
deratingwillbeconstantatacertainpowerperdegreeKelvin(xW/K,seetechnicalspecications),downtothe
steady power which is rated for the typical ambient temperature of 40°C (104°F) and further down.
If the device is supplied with less power than the rated steady power at a certain ambient temperature, the derating
won’t necessarily impact the operation. However, the internal power reduction is always imminent. For example, if
you would run a model with rated 600 W power for a while at 400 W constant actual power, while the power limit
is set to 600 W, and your source would make a voltage step or the load a current step, the power limit of 600 W
could still not be achieved.