On the AC lter enclosure, below the AC connection terminal, there is a
xture to be used in connection with the included cable tie to achieve a
strain relief for the AC cable. See the depiction to the right.
Figure 11 - Strain relief example
2.3.4.5 Grounding the enclosure
For reasons of safety for people working with the device which, amongst other measures, is
achieved by keeping the leakage current as low as possible, the enclosure can be grounded
additionally on this extra grounding point, as depicted on the gure to the right (also marked
as “10” in
“1.8.4. Views”
). It requires a separate protective earth line (PE) and the cross section
of that line must be at least the same as of the ground conductor in the AC supply cable (see
“2.3.4.2. Cross section
).
Figure 12 - Grounding point