EL 9000 B HP Series
3.3 Alarm conditions
This section only gives an overview about device alarms. What to do in case your device in-
dicates an alarm condition is described in section „3.6 Alarms and monitoring“ on page 54.
Asabasicprinciple,allalarmconditionsaresignalledoptically(text+messageinthedisplay),acoustically(if
activated) and as a readable status and alarm counter via the digital interface. In addition, the alarms OT and
OVP are reported as signals on the analogue interface. For later acquisition, an alarm counter can be read from
the display or via digital interface.
3.3.1 Power Fail
Power Fail (PF) indicates an alarm condition which may have various causes:
• AC input voltage too low (mains undervoltage, mains failure)
As soon as a power fail occurs, the device will stop to sink power and switch off the DC input. In case the power fail
was an undervoltage and is gone later on, the alarm will vanish from display and doesn’t require to be acknowledged.
The condition of the DC input after a gone PF alarm can be determined in the MENU. See 3.4.3.
Switching off the device with the mains switch can not be distinguished from a mains blackout
and thus the device will signalise a PF alarm every time the device is switched off. This can
be ignored.
3.3.2 Overtemperature
An overtemperature alarm (OT) can occur due to an excess temperature inside the device and causes it to stop
sinking power temporarily. After cooling down, the device will automatically continue to supply power, while the
condition of the DC input remains and the alarm doesn’t require to be acknowledged.
3.3.3 Overvoltage
An overvoltage alarm (OVP) will switch off the DC input and can occur if:
• the connected voltage source provides a higher voltage to the DC input than set in the overvoltage alarm thresh-
old (OVP, 0...103% U
NOM
)
This function serves to warn the user of the electronic load acoustically or optically that the connected voltage
source has generated an excessive voltage and thereby could damage or even destroy the input circuit and other
parts of the device.
Thedeviceisnotttedwithprotectionfromexternalovervoltage.
3.3.4 Overcurrent
An overcurrent alarm (OCP) will switch off the DC input and can occur if:
• The input current in the DC input exceeds the adjusted OCP limit.
This function serves to protect the voltage and current source so that this is not overloaded and possibly damaged,
rather than offering protection to the electronic load.
3.3.5 Overpower
An overpower alarm (OPP) will switch off the DC input and can occur if:
• the product of the input voltage and input current in the DC input exceeds the adjusted OPP limit.
This function serves to protect the voltage and current source so that this is not overloaded and possibly damaged,
rather than offering protection to the electronic load.