User manual
33
© 2009, Elektro-Automatik GmbH & Co. KG
EN
About the power supply
3. Device description
3.1 Front view
Description of the knobs, buttons and terminals:
1) Power output, safety sockets, poled
The sockets can be used to plug 4mm Bueschel plugs or
to clamp spade lugs.
2) Remote sense input, poled
The remote sense leads are connected here with correct
polarity. For details about the remote sense feature refer to
section „10.1 Remote sense“.
3) Analogue interface, 15pole, D-Sub, female
The socket can be used to remotely control and monitor
the device by means of analogue resp. digital signals. For
more information refer to section „9. Internal analogue
interface“.
4) Pushbutton „Standby“
Is used to switch the device into standby and back to normal
operation.
5) Rotary encoder, right, no stop
Is used to adjust the set value of the output current, the output
power (models from 1kW) or the internal resistance (optional,
only models with unlocked internal resistance control).
Approximately 5 complete turns correspond to 0...100%.
In the setup, it is used to adjust settings.
Also see sections „6.6 Adjusting set values“ and „7. Device
setup“.
6) Rotary encoder, left, no stop
Is used to adjust the set value for the output voltage.
Approximately 5 complete turns correspond to 0...100%.
In the setup, it is used to select parameters.
Also see sections „6.6 Adjusting set values“ and „7. Device
setup“.
7) Control panel and display unit
Figure 1