Instructions

Page 45
EA Elektro-Automatik GmbH
Helmholtzstr. 31-33 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
PS 9000 1U Series
3.8 Loading and saving a user prole
The menu “Proles” serves to select between a default prole and up to 5 user proles. A prole is a collection of
all settings and set values. Upon delivery, or after a reset, all 6 proles have the same settings and all set values
are 0. If the user changes settings or sets target values then these create a work prole which can be saved to
one of the 5 user proles. These proles or the default one can then be switched. The default prole is read-only.
The purpose of a prole is to load a set of set values, settings limits and monitoring thresholds quickly without
having to readjust these. As all HMI settings are saved in the prole, including language, a prole change can also
be accompanied by a change in HMI language.
On calling up the menu page and selecting a prole the most important settings can be seen, but not changed.
How to save the current values and settings (work prole ) as a user prole
1. Switch off the DC output and push button to call the
setup menu.
2. In the menu navigate to “Proles” and push .
3. In the submenu (see gure to the right) select a user prole (1-5) to save to and push again.
4. From the selection on screen chose “Save settings into Prole nand overwrite that prole with the current
settings and values by conrming with .
How to load a user prole
1. Switch off the DC output and push button to call the
setup menu.
2. In the menu navigate to “Proles” and push .
3. In the submenu (see gure to the right) select a user prole (1-5) to load and push again.
4. In the screen you can now select “View Prole n” in order to check the stored settings and to decide, whether
this prole is going to be loaded or not. Navigate to “Load Prole nand conrm with to nally load
the prole into the work prole.