3BUS 208 000 R1301 247Service Workstation Activities
Displaying Reports
The Service Workstation can be used to view Standardize, Check Sample and
Calibrate Sample reports. These reports contain the same information sent to the
Host computer system but the data is displayed differently.
Standardize Report
There are two report arrays associated with the Standardize results:
stdzResultsPending and stdzResults. stdzResultsPending reports on the results of
the last standardize whether the results are within limits or not. stdzResults reports
the results of the last acceptable standardize.
The procedure for displaying these reports is as follows:
1. Access the inspect utility.
2. Type: betaSensor01->stdzResultsPending (for example)
The system will report back the address of that report.
3. Type in the address of the report.
The system will display the results of the last standardize. See Figure A-2.
Figure A-2 Example of Standardize Report
0x307bb2: ”0X307bb2” (STRUCT_PRIVATE)
ISA: BWStdzRslts (STRUCT_SHARED *)
mySelf: 0x307bb2 (long)
problemList 0x30d72 (id)
tempCompFlag 0 ’\0’ (char)
zeroScaleInput 0.000338776 (double)
zeroScaleInputSigma 7.21414e-09 (double)
lastCleanStdz: 0x307bd0 (char [20])
dirt: 0 (double)
fullScaleTempRef: 55.0098 (double)
fullScaleTempSigma: 1.2531e-09 (double)
logRangeClean: 1.45815 (double)
logVoltageRange: 1.45815 (double)
voltageRange 4.29802 (double)
voltageSigma: 3.70175e-05 (double)
tClean 0 (long)
sourceTemperature 54.8 (double)
detectorTemperature 55.06 (double)
sourcePurge 3.70309 (double)
detectorPurge 3.41578 (double)
I> betaSensor01->stdzResults
stdzResults: 0x307bb2 (int *)
I> 0x307662
0x307bb2 (id)