3BUS 208 000 R1301164 Control Panel Operation
Calibrate Sample (Sample Cut)
While the CALIBRATE SAMPLE push button is pressed, the Scanning Platform
begins a special accumulation of measurement single point or profile data. When
the push button is released, this data is forwarded to the Host computer where a
Calibrate Sample report is printed.
The CALIBRATE SAMPLE push button LED is on while the measurement data is
being accumulated, and while the data is being sent to the Host computer.
The CALIBRATE SAMPLE push button functions only when the Scanning
Platform is measuring in the Remote/On Sheet mode and the sensor is in the
measurement mode, either scanning or in single point.
Sample Check
The SAMPLE CHECK push button requests the sensors to take a sample check
measurement. The push button LED is on while the measurement is being taken.
The SAMPLE CHECK push button operates only if the Scanning Platform is in the
Remote/Off-Sheet mode. No report will be sent unless the Sample Check mode is
enabled at the Host computer. However, a local report is available through the
Sensor Health Report.
Standardize
Pressing the STANDARDIZE push button requests the sensors to perform a
standardize. The standardize will be either CLEAN or DIRTY depending on the
setting of the localStdzType instance variable in the fhi01 object. The push button
LED is lit during the standardize cycle. If any of the sensors fail the standardize,
the LED flashed until the sensors all pass standardize. Standardize reports are
available through the HOST display, or through the HEALTH REPORTS at the
Service Workstation.
The sensors will do a CLEAN or DIRT standardize depending on the state of the
defaultStdzType flag associated with each sensor object. If the flag is a “0”, the
standardize will be a DIRT one. If the flag is a “1”, then the standardize will be a
CLEAN one. The only sensors affected by a CLEAN or DIRT request are Basis
Weight, Caliper, and OptiPak. For the other sensors, this discussion is irrelevant.
They will do their normal standardize no matter what the flags are set to.
It is possible to temporarily change the Control Panel default standardize mode from
DIRT to CLEAN by setting the localStdzMode flag in the fhi01 object to a “1”
This assignment is set back to DIRT (“0”) automatically at the next restart.