3BUS 208 000 R1301 151Measurement Setup
5. Adjust the value of dBWdT using the following guidelines:
•If the STANDARD WT value is zero, leave dBWdT as is.
•If the STANDARD WT value is greater than +0.2, decrease the magnitude of
dBWdT.
•If the STANDARD WT value is less than - 0.2, increase the magnitude of
dBWdT.
Note: The range of dBWdT is -20 to 0. A more negative value of dBWdT
means more temperature compensation. A less negative value of
dBWdT means less temperature compensation.
6. Enter the new dBWdT and repeat the sample checks with the heat source On.
The STANDARD WT should be 0.00
± 0.2.
If the ATC operation does not meet specifications, check the thermistor
operation.