3BUS 208 000 R1301 145Measurement Setup
Gloss Sensor
Table 6-27 Standardize Limits for the Gloss Sensor
Host Computer
Reporting Name
Variable Name In
Gloss Health
Report
Variable Description
Acceptable
Range
stdzAdjNominal Nominal value for limit
checking the maximum signal
output on air. It should be about
1.7 VDC.
1.71 VDC
stdzAdjRange Range limit of the maximum
signal output on air. It should
be set to 1 VDC.
± 1 VDC
stdzZeroOffset-
Nominal
Nominal value for limit
checking the minimum signal
output with the measurement
clamp. It should be about 0.06
VDC.
0.26 VDC
stdzZeroOffset-Range Range limit of the minimum
signal output with the
measurement clamped. It
should be set to 0.1 VDC.
± 11 VDC
stdzVacuumNominal Nominal value for limit
checking the head pressure
associated with the vacuum. It
should be about 7.5 VDC. Note
that increasing pressure
decreases the voltage.
7.5 VDC
stdzVacuumRange Range limit of the head pressure
associated with the vacuum. It
should be set to 1.5 VDC.
± 1.5 VDC