Manual

Appendix G Device Diagnostic C O D E S
150 PowerSmart+ Advanced Power Quality Meter
Appendix G Device Diagnostic Codes
Diagnostic
Code
Description Reason
2 Memory/Data fault Hardware failure
3 Hardware watchdog
reset
Hardware failure
5 CPU exception Hardware failure
6 Run-time software error Hardware failure
7 Software watchdog
timeout
Hardware failure
8 Power Down/Up Normal power-up sequence
9 Warm restart External restart via
communications or by firmware
upgrade
10 Configuration reset Corrupted setup data has been
replaced with the default
configuration
11 RTC fault The clock time has been lost
13 Low battery Battery replacement is required.
With auto-reset.
15 EEPROM fault Hardware failure
See Device Diagnostics in Chapter 4 for more information on the PowerSmart+ Advanced
PQM built-in diagnostics.