Specifications
TTroubleshooting
Incorrect control mode of the piezo channel
⇒ Check the current control mode using the ONL? command or on the
main screen of the E-517 display. The current control mode of a piezo
channel determines the applicable control sources for the output voltage
and hence for the axis motion. See
"Control Modes" (p. 23) for details.
Incorrect configuration
⇒ Check the parameter settings on the E-517 with the SPA? and SEP?
commands.
Incorrect GCS syntax version
⇒ Check the current GCS syntax version using the CSV? command or on
the main screen of the E-517 display ("E517" or "E516", in the top right
corner of the display). The current GCS syntax version determines the
applicable GCS commands and software tools (except for the PITerminal,
the software on the E-517 CD can not be used when the E-517 is switched
to the E-516 GCS syntax version). See
"Select Command Set Version"
(p.
41) for details.
The high voltage output of the piezo control electronics is
deactivated, while the communication with the E-517 is still
possible
⇒ The high voltage output of the piezo control electronics may be
deactivated automatically when an internal temperature sensor detects
overheating. To reactivate the high voltage output, let the device cool down
and power-cycle or reboot the E-517.
Note that the wave generator output will continue even if the high voltage
output is deactivated, i.e. if a certain number of output cycles was set, they
may have already finished when the high voltage output is reactivated.
How to avoid overheating:
Keep the ambient temperature at a noncritical value: Note that the
difference between ambient temperature and internal temperature of the
piezo control electronics normally is about 20 Centigrade (36 Fahrenheit)
degrees.
Place the system in a location with adequate ventilation. Allow at least
10 cm (4 inches) clearance from the top and the rear of the unit and 5 cm
(2 inches) from each side. If this is not possible, keep the ambient
temperature low. Never cover the ventilation openings of the piezo control
electronics as this will impede ventilation.