Specifications

TTroubleshooting
The first time you connect over the USB interface, be sure you are
logged on the PC as a user having administrator rights. After the E-517 is
powered on, a message will appear on the PC screen saying that new
hardware has been detected. Follow the on-screen instructions and insert
the E-517 CD. The required FTDI hardware drivers are found in the
\USB_Serial_Driver directory.
Controller was power-cycled or rebooted
With TCP/IP and USB connections, communication can not be
maintained after the E-517 is power-cycled or rebooted. The connection
must then be closed and reopened.
Another program is using the interface
Close the other program.
Specific software has problems
See if the system works with some other software, e.g. a terminal or
development environment. You can, for example, test the communication
by simply starting a terminal program, e.g. PI Terminal, and entering
commands like *IDN? or HLP?. Note that multi-character commands are
transferred as terminated by a LF (line feed) character and are executed
only after the LF is received.
Stage does not move
Cable not connected properly
Check the connecting cable(s)
Stage or stage cable is defective
Exchange stage with a working stage to test a new combination of
controller and stage:
With E-7xx controllers, this is only possible with stages which are equipped
with ID-chips.
With E-861 controllers, the encoder hardware of closed-loop systems must
www.pi.ws E-517 PZ214Equ Release 1.0.0 Page 64