Specifications

Communication
4.3 TCP/IP Connection
The TCP/IP connection is available on the E-517 front panel of the
controller, via the RJ 45 socket with the network icon. The access differs
depending on the network availability, which can be as follows:
Network with DHCP server
PC equipped with an Ethernet connection or network without DHCP
server
For the default IP address, IP mask and startup behaviour settings of the
TCP/IP interface see
"Default and Current Settings" (p. 46).
!
CAUTION
With TCP/IP and USB connections, communication can not be
maintained after the E-517 is power-cycled or rebooted. The
connection must then be closed and reopened.
NOTES
Make sure that your network administrator has not set the network to
forbid unknown devices like the E-517 from logging on.
Only one TCP/IP connection is allowed at any one time: Presently, only
one port (50000) is available at the E-517 so that only one application
at a time can use the TCP/IP connection. The default port setting can
not be changed.
4.3.1 Network with DHCP Server
NOTES
For successful connection, the E-517 startup behaviour for IP address
configuration must be set to "Use DHCP to obtain IP address"
(IPSTART=1; default selection).
You can check the IPSTART settings with the IFS? command
(power-on default setting) and the IFC? command (current setting).
If the IPSTART setting is "Use IP address defined with IPADR"
(IPSTART=0), send
IFS 100 IPSTART 1
The change becomes active with the next power-on or reboot.
www.pi.ws E-517 PZ214Equ Release 1.0.0 Page 48