Macro execution can be stopped from the command line with #24. The STP
command stops macro execution only when it is part of the macro.
A running macro may not be deleted.
You can query with #8 if a macro is currently running on the controller.
3.8.3 Start-Up Macro
With MAC DEF <macroname> it is possible to set the specified macro as
start-up macro. This macro will be automatically executed with the next
power-on or reboot of the controller.
Example:
MAC BEG init
ONL 1 1 2 1 3 1
SVO A 1 B 1 C 1
VCO A 1 B 1 C 1
DCO A 1 B 1 C 1
VEL A 20 B 20 C 20
MAC END
This macro switches the E-517 to ONLINE mode and closed-loop
operation, enables velocity control mode and drift compensation mode and
sets the velocity to 20 µm/s.
To ask for the current start-up macro setting, send
MAC DEF?
To undo the current start-up macro selection, send
MAC DEF
i.e. omit <macroname>.
Deleting the current start-up macro with MAC DEL <macroname> also
deletes the start-up macro selection.
NOTE
To skip the execution of the start-up macro, press the trackball once
immediately after you have powered on or rebooted the piezo control
electronics (while "Physik Instrumente" is still shown on the display).
For details about the trackball handling, see "Trackball Functions" (p.
22).