Specifications

Operation
8 To analyze the recorded data and display it as a Bode frequency
response diagram, click on the
FFT button in the icon bar above
the graphics display.
9 Optionally, you can configure the display properties and activate
cursors using the buttons in the icon bar above the graphics
display (see the PIMikroMoveā„¢ manual for details).
This helps to identify the resonance peak on the Bode plot. For
example, you can place a cursor on the peak and read out the
cursor value which is displayed beside the graph (see figure below;
the resonant frequency is about 824 Hz).
10 When the measurement is finished, open the
Command Entry
window again and set the Digital Filter Type parameter back to its
initial value (see step 3).
NOTE
If the resonant frequency should not be damped well enough (the
measured resonance peak is higher than the highest amplitude value
for lower frequencies), then you should adjust the notch filter (center
frequency, damping) of the analog piezo control electronics as
described in the User manual of the E-802 submodule.