Specifications

Operation
Send: SPA 1 0x05000000 0
to deactivate the digital filter.
Close the
Command Entry window.
4 Open the Data Recorder window via the E-517 Show/Hide data
recorder ...
menu entry.
5 Configure recording:
Press the
Configure... button in the down right corner of the Data
Recorder
window. In the Configure Data Recorder dialog which
opens, check the box in the
Read column and select "Current
position of axis" in the
Option column for the axis to be measured.
In the example shown in the figure below, the current position of
axis A will be recorded and read afterwards.
Click
OK to close the dialog.
6 Configure the measurement in the bottom line of the
Data
Recorder
window:
Select the axis to be measured in the leftmost field (A in this
example).
Deselect the
Servo checkbox since the measurement must be
done in open-loop operation.
Enter the amplitude of the impulse in V in the field between the
start buttons for step and impulse response measurements and
press Enter on your keyboard:
The Record Rate value should be 1.
7 Start the impulse response measurement by clicking on the button
right beside the amplitude field. An impulse move is commanded
(with the IMP command) and the recorded position is displayed in
the graphics pane of the
Data Recorder window.
www.pi.ws E-517 PZ214Equ Release 1.0.0 Page 33