The data recorder sampling period can be read with the RTR? command.
The answer gives the value of the Data Recorder Table Rate parameter (ID
0x16000000) whose default value is one servo cycle. You can cover longer
periods by increasing this value. Use the RTR command or change the
parameter value directly.
The current data recorder configuration made with DRC and RTR is saved
with the WPA command, in addition to the current parameter values of the
E-517 and other settings (see the WPA description for details). When the
controller is powered down, the contents of the data recorder tables and all
settings which were not saved with WPA are lost.
3.5.2 Application Example: Frequency Response
Measurements
Mechanical resonances of the system may exaggerate the response to
certain frequencies in the control signal. To determine the resonant
frequencies of the mechanics, use the E-517 data recorder to observe the
system response to an impulse in open-loop operation.
NOTE
Very stiff mechanics, e.g. S-325, S-316, P-8xx or P-2xx, have resonant
frequencies of several kilohertz. With such mechanics, the frequency
response measurement described below will not deliver meaningful
results since the sensor bandwidth of the system is limited by the
hardware settings on the E-509 analog sensor/servo module.
Meaningful results can only be attained with mechanics whose
resonant frequency is 1 kHz.
Proceed as follows for the axis whose resonant frequency is to be
measured. In this example, the resonant frequency of axis A is to be
measured:
1 Make sure the mechanics is mounted and connected to the piezo
control electronics in exactly the same way as in the application.
The load on the mechanics is especially important.
2 Start PIMikroMove™ on a host PC connected to the E-517 (see the
PIMikroMove™ software manual on the included CD for details).
3 In the
Command Entry window of PIMikroMove™, deactivate the
digital filtering of the sensor signal for the sensor channel which
belongs to the measured axis. This is recommended to achieve
more reliable results. In this example, you have to do this for the
first channel:
Send: SPA? 1 0x05000000
to query the current filter type setting, response should be 2 (= FIR
filter; default setting of the Digital Filter Type parameter)