Specifications
Operation
3.1.2 Display Screens
The individual display screens can be accessed by the trackball (p. 22) on
the E-517 front panel. You can check the different screens by powering up
the piezo control electronics with no actuator/stage and no host PC
connected.
The following screens are available via the trackball:
■ "Main screen", the first line shows the current control mode ("o") and
servo mode ("s") for all channels (0 = OFFLINE / open-loop
operation, I = ONLINE / closed-loop operation; see
"Modes of
Operation"
(p. 23) for details) and the active syntax version of the PI
General Command Set (E517 or E516, see
"Select Command Set
Version"
(p. 41) for details). Lines 2 to 4 show the current voltage and
position values (in V and μm) for all connected channels. The current
voltage value can also be read with the VOL? command and the
current position value with the TSP? command.
The main screen is active upon power-on and reboot.
■ DISPLAY SERVICE, where you can adjust the brightness and the
contrast of the display. Ranges are 10-31 for brightness and 5-31 for
contrast. The settings can be permanently saved as power-on
defaults with WPA 100 (see WPA command) and are also available
via controller parameters (brightness: ID 0x0E000D00; contrast: ID
0x0E000D01; see "Controller Parameters" in the E-517 User Manual
for more information)
■ COMMUNICATION, where you can select the RS-232 baud rate and
the IEEE 488 address. The settings can be permanently saved as
power-on defaults with WPA 100 (see WPA command). They are
also available with IFC, IFC? (temporary settings) and IFS, IFS?
(default settings) and via controller parameters (RS-232 baud rate: ID
0x11000400; IEEE 488 address: ID 0x11000900; see
"Default and
Current Settings"
(p. 46) and "Controller Parameters" in the E-517
User Manual for more information)
www.pi.ws E-517 PZ214Equ Release 1.0.0 Page 21