Specifications
First Steps
5 Give the E-517 interface / display module control over the output
voltage channels of the piezo control electronics by switching them
to ONLINE mode. This can be done in the
Start up controller
window which should open automatically with the
Start up E-517
step, see figure below. A channel is in ONLINE mode when its
check box is checked, otherwise it is in OFFLINE mode. When the
mode selection is finished, press the
OK button. The
PIMikroMove™ main window will open.
You can reopen the window for ONLINE / OFFLINE selection at
any time via the E-517 ⇒ Start up axes ... menu sequence from
the PIMikroMove™ main window.
Notes:
In ONLINE mode the SERVO switches of all channels must be set
to OFF on the piezo control electronics. This gives the E-517
complete control over the servo mode settings.
Channels which are in OFFLINE mode can not be commanded by
the E-517 (and hence not by the host PC) but only via controls on
the piezo control electronics (e.g. analog control input and / or
DC-offset potentiometers).
The current mode of the channels is also visible in the main screen
of the E-517 display (top left corner; 'I' for ONLINE, 'O' for
OFFLINE).
www.pi.ws E-517 PZ214Equ Release 1.0.0 Page 15