Specifications

First Steps
program.
Proceed as follows to create a parameter file:
1 Power on the system. Find details in a
dedicated section (p. 31).
2 Start PIMikroMove™ on the host PC and establish a connection to
the E-517 as described in
"Quick Start" (p. 13).
3 In the PIMikroMove™ main window, open the Device Parameter
Configuration window using the E-517 Parameter Configuration
menu sequence.
4 In the Device Parameter Configuration window, save the controller
parameters to a file. Use the
Save or Save As buttons in the top
left-hand corner of the window, or use the File Save Edit Values
or File Save Edit Values As menu sequences. In fact, the values
from the
Edit Mask column are saved.
In the Device Parameter Configuration window of PIMikroMove™, proceed
as follows to load back the content of a parameter file (with the extension
.pam):
1 Use the File Load and select menu sequence. The loaded
parameter values are written in the corresponding
Edit Mask fields.
2 You can copy the loaded parameter values from the
Edit Mask
fields to the RAM (volatile memory) or non-volatile memory of your
controller using the three
Write selected edit values... buttons in the
top right-hand corner of the window.
Note that depending on the controller, changing parameters may
require a certain command control level (CCL). You can switch to
command level 1 and change level-1 parameters (see
CCL
column), while command levels >1 are reserved for service
personnel. In the dialog which pops up automatically when you try
www.pi.ws E-517 PZ214Equ Release 1.0.0 Page 12