Datasheet

v1.0 / Modification rights reserved
14
EEH210
The DA_EN bit enables the DA signal on pin 9. Normally inactive, the DA output signal becomes active on new data
available: logical OR of the bits STATUS_REG[1] and STAUTS_REG[0] for humidity and temperature, respectively. The
DA signal returns inactive after both HUMIDITY_OUT_H and TEMP_OUT_H registers are read.
STATUS_REG
Table 21: STATUS_REG
7 6 5 4 3 2 1 0
Reserved H_DA T_DA
Address: 27h (R)
Description: Status register; the content of this register is updated every one-shot reading, and after completion of
every ODR cycle, regardless of BDU value in CTRL_REG1.
[7:2] Reserved
[1] H_DA: Humidity data available
(0: new data for Humidity is not yet available, 1: new data for Humidity is available)
[0] T_DA: Temperature data available
(0: new data for temperature is not yet available, 1: new data for temperature is available)
H_DA is set to 1 whenever a new humidity sample is available. H_DA is cleared anytime HUMIDITY_OUT_H (29h) reg-
ister is read.
T_DA is set to 1 whenever a new temperature sample is available. T_DA is cleared anytime TEMP_OUT_H (2Bh) register
is read.
HUMIDITY_OUT_L
Table 22: HUMIDITY_OUT_L
7 6 5 4 3 2 1 0
HOUT7 HOUT6 HOUT5 HOUT4 HOUT3 HOUT2 HOUT1 HOUT0
Address: 28h (R)
Description: Humidity data
[7:0] HOUT7-HOUT0: Humidity data LSB
HUMIDITY_OUT_H
Table 23: HUMIDITY_OUT_H
15 14 13 12 11 10 9 8
HOUT15 HOUT14 HOUT13 HOUT12 HOUT11 HOUT10 HOUT9 HOUT8
Address: 29h (R)
Description: Humidity data
[7:0] HOUT15-HOUT8: Humidity data MSB
TEMP_OUT_L
Table 24: TEMP_OUT_L
7 6 5 4 3 2 1 0
TOUT7 TOUT6 TOUT5 TOUT4 TOUT3 TOUT2 TOUT1 TOUT0
Address: 2Ah (R)