Datasheet

61
EE471
v1.3 / Modification rights reserved
Dimensions in mm (inch)
80.8 (3.18)
55.3 (2.18)
39.5 (1.56)
Ø4
(0.16)
63 (2.48)
42 (1.65)
55 (2.17)
5
(0.2)
Ø6
(0.24)
star
pressing
55.3 (2.18)
EE471-T6xx
Output: 4...20 mA
Output: 0...10 V
power supply
15...35 V DC
24 V AC ±20%
power supply
20...35 V DC R
L
< 500 Ω
11...35 V DC R
L
< 50 Ω
EE471-T3xx
Active Output
Passive Output
EE471-Txx
Connection Diagram
General
Insulation resistance (remote probe) > 100 MΩ at 20 °C (68 °F)
Response time τ
63
< 1 min, at 3 m/s (590 ft/min) air velocity
< 30 s, with immersion well in liquid water bath
Sensor sleeve material stainless steel (1.4571 / 316Ti)
Cable material PVC
Enclosure material polycarbonate, UL94-V0 approved
Protection class IP65 / NEMA 4 (enclosure), IP67 / NEMA 4 (remote probe)
Cable gland M16x1.5, UL94-V2
Storage temperature -30 °C…+70 °C
(-22 °F...+158 °F)
Working and storage humidity range 5 % rh...95 % rh, no condensation
Additional Information
Wire Resistance / Temperature Offset
(Only relevant for passive output!)
Cable length Wire resistance Temperature offset for Pt100
*)
0.5 m (1.64 ft) 0.124 0.32 °C (32.576 °F)
2 m (6.56 ft) 0.364 0.93 °C (33.674 °F)
3 m (9.84 ft) 0.520 1.33 °C (34.394 °F)
*) For high-resistance T-sensors (R ≥ 1000 Ω) the temperature offset is negligible.
- EE471 Temperature sensor according ordering guide
- Cable gland
- Two self-adhesive labels for configuration changes (see user guide at www.epluse.com/relabeling)
- Test report according to DIN EN10204 - 2.2 (for active output only)
Scope of Supply