Datasheet
30
EE220
v1.0
EE220 Series
Humidity / Temperature Transmitter
with interchangeable probes
Unique for the EE220 series are the interchangeable sensing
probes.
The calibration data is stored in the probes, therefore a probe
replacement does not affect the accuracy of EE220.
The outstanding accuracy over the entire temperature range is
based on very precise calibration methods and on the latest
microprocessor technology. Well-proven E+E humidity sensor
elements ensure excellent long-term stability.
For high temperature applications (up to +80°C / +176°F) or in case
of limited space availability, the sensing probes can be connected
to EE220 housing with cables (2m, 5m or 10m / 6.6ft, 16.4ft or
32.8ft) without any repercussions for the overall accuracy of the
instrument.
Voltage 0 - 1 / 10V or current 4 - 20mA (2 wire) EE220-outputs
are available, of which the temperature output can be scaled
according to the application (see ordering guide).
EE220 is suitable for direct wall mounting and for installation on
rails according to DIN EN 50022.
The optional display indicates the actual RH- and T-values.
Duct mounting can be done easily with the optional
duct mounting kit.
pharmaceutical industry
clean rooms
storage rooms
green houses
cooling chambers
interchangeable probes
remote sensing probe up to 10m (32.8ft)
measuring range 0...100% RH / -40...80°C (-40...176°F)
optional display
easy field loop-calibration
Typical Applications
Features
RH probe
T probe
Field calibration of humidity and temperature
In the pharmaceutical and biotechnology industry a Loop-
Calibration of the RH- and T-outputs, recommended by the
FDA (Food and Drug Administration), can easily be performed
utilizing separate RH- and T-probes (Type: EE220-xxx2x).
The RH- and T-outputs can be adjusted with push buttons on
the printed circuit board.
EE220 + EE07
in polycarbonate
housing
Reference probes
As useful accessories reference probes (incl. test report) representing
fixed humidity and temperature values are available.
They shall be installed instead of the measuring probes to check
function and accuracy of the evaluation unit.
One probe simulates high humidity and low temperature, the other
low humidity and high temperature, to check the upper and lower end
of both analogue outputs.
EE220 + EE07
in metal housing