Datasheet
35
EE210
v1.7 / Modification rights reserved
Measured Values
Relative Humidity
Sensor E+E Sensor HCT01-00D
Analog output 0...100% RH 0-5 V -1 mA < I
L
< 1 mA
0-10 V -1 mA < I
L
< 1 mA
4-20 mA (two-wire) R
L
≤ 500 Ohm
Working range 0...100% RH
RH accuracy (incl. hysteresis, non-linearity and repeatability)
Wall & duct version:
-15...40°C (5...104°F) ≤90% RH ±(1.3 + 0.003*measured value) % RH
-15...40°C (5...104°F) >90% RH ± 2.3% RH
-40...60°C (0...140°F) ±(1.5 + 0.015*measured value) % RH
Remote probe version
at 20°C (68°F) ±2.5% RH
Temperature
Sensor Pt1000 (tolerance class B, DIN EN 60751) integrated in HCT01
Analog output
1)
0-5 V -1 mA < I
L
< 1 mA
0-10 V -1 mA < I
L
< 1 mA
4-20 mA R
L
≤ 500 Ohm
T-accuracy wall & duct remote probe
General
Power supply
for 0-5 V / 0-10 V 15 - 35V DC
2)
or 24V AC ±20%
for 4-20 mA 10V + R
L
x 20 mA < V+ < 30V DC
Current consumption
Voltage output DC supply typ. 3.3mA; with display typ. 3.6mA
AC supply typ. 34mA; with display typ. 37mA
Current output DC supply max. 40mA
Digital interface DC supply typ. 5mA; with display typ. 19mA
AC supply typ. 52mA; with display typ. 118mA
Connection Screw terminals, max. 1.5 mm
2
Housing material Polycarbonate, UL94V-0 (with Display UL94HB) approved
Protection class IP65
Cable gland M16 x 1,5
Probe cable (type C) PVC, Ø 4.3mm, 4 x 0.25 mm
2
Sensor protection E+E Coating
Electromagnetic compatibility EN61326-1 EN61326-2-3
Industrial Environment
Temperature ranges Operating temperature: -40...60°C (-40...140°F)
Storage temperature: -40...60°C (-40...140°F)
Temperature ranges with display Operating temperature: -20...50°C (-4...122°F)
Storage temperature: -20...60°C (-4...140°F)
Technical Data
1)
Output scaling see Ordering Guide
2)
USA & Canada: class 2 supply required, max. supply voltage 30V
∆ °C
°C
∆ °C
°C