Datasheet

41
EE160
v1.4 / Modification rights reserved
Measured values
Relative Humidity
Sensor E+E Sensor HCT01-00D
Analog output 0...100% RH 0-10 V -1 mA < I
L
< 1 mA oder
4-20 mA (two-wire) R
L
< 500 Ohm
Digital output RS485
Working range 10...95% RH
Accuracy at 20°C ±2.5% RH
Temperature dependency typ. ±0.03% RHC
Temperature
Sensor Pt1000 (tolerance class B, DIN EN 60751)
Analog output
1)
0-10 V
4-20 mA
Digital output RS485
T-Accuracy at 20°C ±0.3°C
passive T-output see ordering code
General
Power supply
for 0 - 10 V / RS485 15 - 35V DC or 24V AC ±20%
for 4 - 20 mA 10V + R
L
x 20 mA < U
V
< 35V DC
Current consumption
Analog with DC power supply typ. 5mA
with AC power supply typ. 13mA
eff
Digital with DC power supply typ. 15mA
with AC power supply typ. 25mA
eff
Connection Screw terminals, max. 1.5 mm
2
Housing / protection class Polycarbonate (UL listed) / IP65
Cable gland M16 x 1.5
Sensor protection membrane filter
Electromagnetic compatibility EN61326-1
EN61326-2-3
Temperature ranges Operating temperature: -15...60°C
( 5...140°F)
Storage temperature: -25...60°C ( -13...140°F)
Technical data
CABLE GLAND
M16x1.5
CONDUIT KNOCKOUT
Type B
Type A
Dimensions (mm)
1)
Output scaling see Ordering Guide